Product features
● Reference accuracy Digital and intelligent: ±0.2% calibration range;
● Stability: Digital and intelligent: 6 months, ±0.2%URL;
●Ambient temperature limits: -40 to 85℃;
● Process temperature limit: -40°C to 104°C.
● Storage temperature limit: -51 to 85℃;
● Ambient humidity: 0-100% relative humidity;
● Effect of ambient temperature: ±0.2%URL/56℃;
● Overall error: ±(0.2%URL+0.18% calibration range)/56℃;
● Zero error: ±0.5%URL/56℃;
● Overall error: ±(0.5%URL+0.5% calibration range) /56℃;
● Influence of static pressure: The zero error is ±0.5%URL;
● Influence of vibration: The vibration effect under 200Hz is ±0.5%URL/g;
● Power supply effect: Less than ±0.005% output range/volt;
● Mounting position influence: The zero drift is no more than ±0.25kPa;
● Measuring range: Differential pressure: 0-0.125~1.5KPa, static pressure: 1, 4 and 6.9MPa;
● Zero and measuring range: Adjustment with local range and zero button or remote adjustment with HART manual actuator;
●Zero positive and negative transfer: At the moment of zero negative transfer, the lower limit of measuring range must be more than or equal to -URL. At the time of zero positive transfer, the upper limit of measuring range must be less than or equal to +URL. The calibration range is no less than the minimum measuring range;
● Output: 4~20mA DC. The user can choose linear or square root output. The digital process variables are superimposed on 4 ~ 20mA DC signal and can be used for the upper computer with HART protocol used;
● Static pressure and overpressure limits: No damage will be incurred by adding 0-6.9MPa pressure to any side of the transmitter. At the moment of working within the static pressure scope of 3.45kPa-6.9MPa, the performance and specification requirement is met, and the volume change is less than 0.16cm3.
Product model selection
PTC-DR | Slight differential pressure transmitter | ||||
Measuring range 0-0.1~1.5 KPa) | |||||
Output E Non-intelligent type (4~20mA) S Intelligent type (Hart protocol+4~mA) J Intelligent square-root output (Hart protocol+4~mA) | |||||
Structure material 22 Flange joint/blow-down valve/isolation diaphragm/filling liquid: 316/316/316L/silicone oil 23 Flange joint/blow-down valve/isolation diaphragm/filling liquid: 316/316/hastelloy C/silicone oil 24 Flange joint/blow-down valve/isolation diaphragm/filling liquid: 316/316/monel/silicone oil 25 Flange joint/blow-down valve/isolation diaphragm/filling liquid: 316/316/tantalum/silicone oil 33 Flange joint/blow-down valve/isolation diaphragm/filling liquid: Hastelloy C/ hastelloy C/ hastelloy C/silicone oil 33 Flange joint/blow-down valve/isolation diaphragm/filling liquid: Hastelloy C/ hastelloy C/ hastelloy C/silicone oil 35 Flange joint/blow-down valve/isolation diaphragm/filling liquid: Hastelloy C/ hastelloy C/tantalum/silicone oil 44 Flange joint/blow-down valve/isolation diaphragm/filling liquid: Monel/monel/monel/silicone oil | |||||
Static pressure A 1Mpa B 4Mpa | |||||
Additional function/optional M1 0-100% linear dial indicator M2 LED displayer meter M3 LED display meter B1 Tube bending mounting plate B2 Plate bending mounting plate B3 Tube flat mounting plate C0 1/2-14NPT taper pipe female adapter C1 1/2-14NPT lead pressure joint back welding connecting tube 4 C2 T-square threaded connector M20X1.5 E Flame-proof type DIIBT4 D Intrinsic safety type IAIICT6 | |||||