GDP online dew-point instrument
View: Pubdate:2022-04-25

Product overview 

The GDP online dew-point instrument is capable of preventing dust, dirt and water splashing based on the structure design and can be used in harsh environments. It's of excellent long-term anti-condensation stability and durability, thus minimal maintenance is required. 

It's an ideal choice for industrial application and mainly used for air and plastic desiccant, drying room, drying gas and high-voltage breaker. 

 In bad conditions (such as comprehensive influence of low humidity and high air temperature), the measurement result is also accurate and reliable. 

 It's mainly used for petrochemical engineering, 

 natural gas, industrial gas, semiconductor industry, drying industry, food industry, power sector, machine manufacturing, air separation industry, pharmaceutical industry, lithium battery new energy industry, etc. 

Product features 

● Many output parameters: Display various humidity parameters, such as dew point/frost point temperature, relative humidity, ppmv(volume of moisture/dry gas volume) and environment temperature; 

● The measurement data is displayed numerically or graphically on the LCD display screen and the output readings can be converted into real-time graphics by a key; 

● The user interface is easily used, and the fixed keys refer to the graph, hold/save and record keys; 

● The structure is light and firm with waterproof and dustproof enclosure protection. It can be used stably in an extremely hostile environment and is suitable for various occasions. 


Technical indicators 

● Measuring performance 


Dew point

Sensor:

GDP-M350

Measuring range:

-25··+100℃(-13···+212°℉)

Accuracy:

±2℃(±3.6°℉)Td/f

Response time: 63[90]

From wet to dry:

5s-15s

From wet to dry:


45s [5min]


Mixing ratio


Measuring range (typical): 0··1000g/kg(0··7000 gr/lbs)

Accuracy:

Reading+-12%

Probe operating temperature scope


0··+350℃(+32···+662°℉)

Operating temperature scope of the body for connection on the probe


-40···+80℃(-40···+176°℉)

Storage temperature

-40···+80℃(-40···+176°℉)

Measuring environment: Air, nitrogen, hydrogen, argon, helium and oxygen

IP protection level

IP66



 

● Input and output

Operating voltage

15··30 VDC

Current consumption


10mA (typical value), 500Ma (maximum value)

Digital output

RS-485, non-isolated

Protocol

Modbus RTU

Output parameters

Dew point temperature, mixing ratio, water concentration, vapour pressure and water mass fraction

● Machinery specification

Connector joint

M125 pointer A standard

Probe cable length




2m


Material

Probe


Stainless steel AISI316L

Probe body

Stainless steel AISI316L

Cable enclosure

Plastic FEP

Cooling device

Stainless steel and aluminum


 

● Measurement parameter

Sensor

GDP-M80 multiparameter sensor


Dew point accuracy

±2℃(±3.6°℉)

Pressure accuracy at 23℃ (73.4°F)

±0.4%FS

Pressure and temperature dependence bar/10℃(18°F)

±0.01

ppm Accuracy (7 bar)

± (12of 14ppmreading)

Dew point

70··+30℃(-94···+86°℉)

Pressure, absolute

1···12 bar(14.5···174 psi)

Temperature (available only when the output RS-485 is chosen)

-40···+80℃(-40···+176℉)

ppm moisture value (based on the volume)

1···40 000 ppm

Dew point, having been converted into atmospheric pressure

-75··+30℃(-103···+86°℉)

  

● Measuring performance


● Input and output

Operating temperature of electron device

40··+60℃(-40···+140°℉)

Accuracy of analog output

±0.01V/±0.01 mA

Operating pressure

1···12 bar(14.5··174 psi)

Digital output RS-485, non-isolated, Modbus RTU protocol

Mechanical temperature resistance

0··50 bar(0··725 psi)

Connector

4-pin M 8

Relative humidity

0··100%

Operating voltage

Measured gas

Air/non-corrosive gas

Current output

21···28 VDC

Sampling gas velocity

No effect on measurement precision

Voltage output and/or use at low temperature (-40··-20℃ (-40···-4°℉))

20···28 VDC

Storage temperature

Only RS-485

15···28 VDC

Only transmitter

-40···+80℃(-40···+176°F)

Analog output (two channels)

Temperature accuracy

Current output

0··20mA4··20mA

0…40℃(+32···+104°℉)

±0.5℃(±0.9°℉)

Voltage output

0··5V,0··10V

0-40···80℃(-40···+176°℉)

±1℃(±1.8°℉)

Input current

Sensor response time

Carrying current in the conventional measurement process

20mAnegative

Pressure response time

≤1s

Carrying current in the self diagnosis process

300mAnegative

Shipping and packaging

-20···+80℃(-4···+176°℉)

External load

Dew point response time: 63[90] under the condition of 20℃ and 1bar

Current output

Maximum: 500Ω

-50

5s [10s]

Voltage output

Minimum: 10kΩ

-10→-50℃ Tdf

10s [2.5min]


Working environment

Temperature

-40···+70℃(-40··+158°F)

Relative humidity

0···100RH (maximum: +20℃/+68°℉)

Pressure

0··50 bar(725 psia)


Measuring environment: Used for air, nitrogen, argon, helium and oxygen1) Not suitable for measurement in the hydrogen or pure carbon dioxide

 

Flow rate of sampled gas

No effect on measurement accuracy

Passing EMC certification

EN61326-1, industrial environment


 

● Measurement parameter

Uncalibrated range

-100···+20℃(-148···+68°℉)Td

Sensor

GDP-M70

Typical response time: 63[90] when the gas temperature is 20℃ (68°℉) and the air pressure is 1bar

Recommended calibration interval

2 years

-20··-80℃Td

0.5 min[7.5 min]

Dew point temperature 1)

-80··-20℃Td

2s[5s]

Measuring range:

-80·-20℃(-112··-4°F)Td

Typical long-term stability

Higher than 2℃ (3.6°F)/year

Accuracy

Volume concentration(Kg/m3)

-80··-40℃(-112··-40°℉): ±2℃(3.6°℉)Td

Measuring range (typical)

0··500 ppm

-40·-20℃(-40-·-4°F): ±3℃(5.4°F)Td

Accuracy under 20 ℃ (68 °F) 1013 mbar: ± (20of 0.2ppmreading)



 


● Input and output




Two analog outputs (definable range) 4…20mA, 0…20mA (three-wire), 0…5V and 0…10V

Digital output

RS-485 (two-wire)



Alarm level of analog signal indication

Selectable by the user


Chemical cleaning function information

5V, 10V, 20mA or LED


Accuracy of analog output

11+-0.01V/+-0.01mA

Operating voltage


RS-485 output

11···28 VDC 1)

Voltage output

15···28 VDC 1)

Current output

21··28VDC

Supply voltage

Conventional measurement

Load current20mAload current

Self-diagnosis period

Maximum 220mA pulse

Mains ripple

Maximum: 0.3V

External load

Voltage output

Minimum: 10kΩ

Current output

Maximum: 500Ω

 

● Machinery specification

Mechanical connection: ISO G1/2", NPT 1/2",UNF 3/4"-16",UNF5/8"-18"

Enclosure material (wet parts)

ISI316L

IP protection level

IP66

Stainless steel mesh filter

Filter body AISI303, screen AISI316L, grade: 18μm

Storage temperature scope

0-40··+80℃(-40···+176°℉)

 

 

Scan the qr codeClose
the qr code